tcGas Default Settings 35 Default Settings Appendix
379
tcGas Default Settings
SvO
2
Default Settings
tcGas Settings Factory Adult Factory Pedi Factory Neo
tcpO
2
High 80 mmHg
tcpO
2
Low 50 mmHg
tcpO
2
Alarms On
tcpCO
2
High 50 mmHg
tcpCO
2
Low 30 mmHg
tcpCO
2
Alarms On
Site Time 4.0 hours
Disable Timer Not Allowed
Heat Switch Off (i.e., after Site Timer
elapsed)
No
Tr a n s d u c e r Te m p. 4 3 . 0 ° C
CO
2
Correction (Severinghaus) On
Metabolism Factor 8 mmHg
TcGas Unit mmHg
Tem per atu re Un it
0
C
tcpO
2
Color Blue
tcpCO
2
Color Green
SvO
2
Settings Factory Adult Factory Pedi Factory Neo
Low Limit 60%
High Limit 80%
Alarms On
Light Intensity On
Color Yellow