Resp Safety Information 8 Monitoring Respiration Rate (Resp)
135
Rate adaptive pacemakers: Implanted pacemakers which can adapt to the Minute Volume may
occasionally react on the Impedance measurement used by patient monitors for the determination of
the Resp value and execute pacing with the maximum programmed rate. Switching off the Resp
measurement can prevent this.