Philips MP60/70/80/90 Manual de usuario Pagina 164

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 414
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 163
10 Monitoring NBP Preparing to Measure NBP
144
CAUTION If you spill liquid onto the equipment or accessories, particularly if there is a chance that it can get
inside the tubing or the MMS, contact your service personnel.
Measurement Limitations
Measurements are impossible with heart rate extremes of less than 40 bpm or greater than 300 bpm, or
if the patient is on a heart-lung machine.
The measurement may be inaccurate or impossible:
if a regular arterial pressure pulse is hard to detect
with cardiac arrhythmias
with excessive and continuous patient movement such as shivering or convulsions
with rapid blood pressure changes
with severe shock or hypothermia that reduces blood flow to the peripheries
with obesity, where a thick layer of fat surrounding a limb dampens the oscillations coming from the
artery
on an edematous extremity.
Measurement Methods
There are three methods of measuring NBP:
Manual - measurement on demand.
Auto - continually repeated measurements (between one and 120 minute adjustable interval).
STAT - rapid series of measurements over a five minute period, then the monitor returns to the
previous mode. Use only on supervised patients.
Reference Method
The NBP measurement reference method can be Auscultatory (manual cuff) or Invasive (intra-
arterial). For further information, see the Application Note on NBP supplied on the monitor
documentation CD-ROM.
In Adult and Pediatric mode to check the current setting, select Main Setup -> Measurements
-> NBP, and check whether the Reference setting is set to Auscultatory or Invasive. This
setting can only be changed in Configuration Mode.
In Neonatal mode, to comply with safety standards, invasive is always used as the reference method.
This setting cannot be changed and is not visible in any operating mode.
Preparing to Measure NBP
1 Connect the cuff to the air tubing.
2 Plug the air tubing into the red NBP connector. Avoid compression or restriction of pressure tubes.
Air must pass unrestricted through the tubing.
Vista de pagina 163
1 2 ... 159 160 161 162 163 164 165 166 167 168 169 ... 413 414

Comentarios a estos manuales

Sin comentarios