Philips MP60/70/80/90 Manual de usuario Pagina 375

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 414
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 374
Measurement Specifications 34 Installation and Specifications
355
Temp
Invasive Pressure Alarm Specifications Range Adjustment Delay
Pressure –40 to 360 mmHg
(–5.0 to 48 kPa)
-40 to 30 mmHg
2 mmHg (0.5 kPa)
> 30 mmHg
5 mmHg (1 kPa)
max. 12
seconds
Pulse 25 to 300 bpm Adult:
1 bpm steps (25 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (25 to 50 bpm)
5 bpm steps (50 to 300 bpm)
Tachycardia Difference to high limit 0 to
50 bpm
5 bpm steps max. 14
seconds
Clamping at 150 to 300
bpm
5 bpm steps
Bradycardia Difference to low limit 0 to
50 bpm
5 bpm steps max. 14
seconds
Clamping at 25 to 100 bpm
5 bpm steps
Temp Performance Specifications
Temp Range –1 to 45 °C (30 to 113 °F)
Resolution
0.1 °C (32.2 °F)
Accuracy
±0.1 °C (±0.2 °F)
Average Time Constant Less than 10 seconds
Alarms Range –1 to 45 °C (30 to 113 °F)
Adjustment
-1 to 35 °C (30 to 95 °F): 0.5 °C (1.0 °F) steps
35 to 45 °C (95 to 113 °F): 0.1 °C (0.2 °F) steps
Temp Alarm Specifications Range Adjustment
Tem p Hi gh/ Low Al a rms
–1 to 45 ºC (30 to 113 ºF) -1 to 35 °C (30 to 95 °F), 0.5 °C (1.0 °F) steps
35 to 45 °C (95 to 113 °F), 0.1 °C (0.2 °F) steps
Vista de pagina 374
1 2 ... 370 371 372 373 374 375 376 377 378 379 380 ... 413 414

Comentarios a estos manuales

Sin comentarios