Philips V26CT Manual de usuario Pagina 373

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 392
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 372
Monitor Maintenance
Maintenance 13-7
Maintenance
Preventative Maintenance
Replace backlight tube
assembly
Cleaning procedures
2.5 years of continuous use
or as needed
Perform the Preventive
Maintenance Procedure for
the M1180A #A01/A02 Wall
Mount Assembly every year.
See “Testing and
Maintaining the CMS” in the
“CMS Service Guide” or
“Testing and Maintaining
the V24 and V26“in the “V24
and V26 Service Guide”.
Full performance checks
including safety checks.
At least every two years,
after any repairs, or as
needed.
See “Testing and
Maintaining the CMS” in the
“CMS Service Guide” or
“Testing and Maintaining
the V24 and V26“in the “V24
and V26 Service Guide”.
Check Calibration of NBP &
Sidestream CO
2
Once a year See chapter 3 of the CMS
Service Guide “Maintaining
the CMS”.
Check Calibration of all
parameters except NBP &
Sidestream CO
2
Every 2 years See chapter 3 of the CMS
Service Guide “Maintaining
the CMS”.
Procedure Frequency Source of Information
Vista de pagina 372
1 2 ... 368 369 370 371 372 373 374 375 376 377 378 ... 391 392

Comentarios a estos manuales

Sin comentarios