Philips V26CT Manual de usuario Pagina 187

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 392
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 186
Alarm Recording
Recording Functions 6-15
Recording Functions
A yellow CO
2
alarm would produce a recording ordered as:
If a red P3 alarm occurred, the recording order would then become:
For alarm recordings with oxyCRG (to document neonatal events without
the Neonatal Event capability), the alarm recording items should be set to
the following:
In the event of a Brady, Apnea and/or Desat alarm the recorder will
document an eight minute oxyCRG recording (six minutes preceding the
event and two minutes following the event).
Channel 4 Respiration
Channel 1 ECG
Channel 2 Pressure 1
Channel 3 Pressure 2
Channel 4 CO
2
alarm
Channel 1 ECG
Channel 2 Pressure 1
Channel 3 P3 alarm
Channel 4 CO
2
alarm
AlRecType oxyCRG
All AlarmRec Off
Change AlarmRec to red for ECG, Resp and
SpO
2
.
Vista de pagina 186
1 2 ... 182 183 184 185 186 187 188 189 190 191 192 ... 391 392

Comentarios a estos manuales

Sin comentarios